Rotary Gas Meters: Reading Pressure Loss Across Three Test Setups

Published on 11.27

Rotary Gas Meters: Reading Pressure Loss Across Three Test Setups

Abstract

Zhang, Su, and Lin compared pressure loss on a prototype gas Roots meter using bell, critical-flow nozzle, and master-meter test setups. They also calculated a theoretical curve based on rotor speed. The pressure reductions reported by the test setups and those from digital sensors formed separate columns in the original tables. This article retains that distinction, walks through the original rotor drawings and comparison plot, and explains which observations support a pressure-loss comparison. The result is a reading of one published laboratory study with explicit boundaries on its use for another meter, gas, and installation.

Background and research question

A rotary gas meter repeatedly traps and transfers volumes of gas as its lobed rotors turn. A differential pressure between the inlet and outlet can be observed while the instrument operates. That differential is relevant when reviewing a metering arrangement because the meter contributes to the pressure requirement of the line. The size of that contribution depends on the conditions of the measurement. A useful test record identifies the flow, the gas and its operating state, and the points between which pressure was observed. It also identifies the method that produced the reported value.
Zhang and colleagues worked with one gas-meter prototype of 50 mm nominal diameter. They examined a rotor geometry and reported pressure measurements from several test arrangements. The case provides a practical opportunity to ask how closely pressure readings agree across arrangements, and how a reader should compare these readings with a calculation. A statement near the end of the paper summarizes the maximum pressure loss. The tables allow the reader to inspect the quantities summarized by that statement and to see the differences between measurement channels.

Rotor geometry and scope

Figure 1 shows one quadrant of the rotor profile studied by the authors. The lettered points identify the joining sections of the profile. The paper describes a sequence of circular arcs and an envelope segment. Figure 2 shows the resulting full profile obtained through horizontal and vertical symmetry. These drawings identify the geometry to which the experiment refers. Their role in the pressure comparison is to define the sample and retain the connection between the calculated rotor motion and the physical meter used on the bench.
The published study concerns a particular rotor pair, body and bench. Its reference to conventional pressure drop is a comparison in the authors' discussion, based on their stated experience with the diameter. The publication offers evidence about the prototype and its observed values under the reported conditions. A comparison of two meters for procurement would also need the same definition of pressure loss, comparable flow points and a statement of the medium and test state.
Figure 1. Quarter rotor profile showing the labelled curve junctions. Source: Zhang, Su and Lin [1], original Figure 1.
Quarter rotor profile showing the labelled curve junctions.
Figure 2. Full rotor profile with the original profile and symmetry axes. Source: Zhang, Su and Lin [1], original Figure 2.
Full rotor profile with the original profile and symmetry axes.

Apparatus and measurement method

The paper describes pressure measurement across the inlet and outlet of the meter. The authors made observations with a bell gas-flow system, a critical-flow Venturi nozzle system and a flow system using a master meter. The digital pressure sensing lines provided another record for the same study. The bell system covered a sequence of nominal points from 0.65 to 100 cubic metres per hour. The authors compared three setups at two common nominal points, 70 and 100 cubic metres per hour. The comparison tables distinguish the value recorded by the setup from the value recorded by the sensor.
Before the meter measurements, the authors compared the pressure channels with a micro differential-pressure instrument in a wind tunnel. Table 1 retains the eight fan settings and three sets of observations. The second digital channel followed the comparison instrument closely in the reported rows. For example, at 15 Hz the readings were 205.3 and 205.0 Pa, respectively, while the first digital channel showed 233.2 Pa. The authors report an average difference of approximately 28.8 Pa between the digital channels and describe the fluctuation around it as approximately 2 Pa. This provides context for interpreting the channel readings. The wind tunnel comparison and the subsequent meter results should each be kept with their own settings and reference.
Fan frequency (Hz)
Reference instrument (Pa)
Digital channel 1 (Pa)
Digital channel 2 (Pa)
Channel difference (Pa)
9.0
70.0
100.7
70.4
30.3
10.5
97.1
127.2
97.4
29.8
12.0
127.5
158.0
128.7
29.3
13.5
163.8
193.2
164.6
28.6
15.0
205.0
233.2
205.3
27.9
16.5
250.3
278.2
251.0
27.2
18.0
301.8
328.1
301.5
26.6
25.0
555.5
590.1
559.3
30.8

Calculation and original comparison curve

The source presents a pipe-loss expression as a reference model for one rotor, with ΔP = 1.28ρv² in Eq. (1). The paper identifies ρ as air density and relates v to rotor tip speed. It prints v = 2πrf × 10⁻³ in Eq. (2), N = V/V_A in Eq. (3), and f = N/4 in Eq. (4). The accompanying text describes four pulses per rotation. Table 2 reproduces the authors' values for the nominal point, run time, cumulative volume, pulse quantity, frequency, tip speed and calculated loss for one rotor.
Equation (2) as printed includes the scaling factor and the adjacent text gives the radius in metres. The units and scaling for the calculation should be clarified with the original authors or source before reusing it for a different meter. The data presented here retain the published expressions and table values. At 100 cubic metres per hour the table lists 193.9085 Pa for one rotor, and at 70 cubic metres per hour it lists 94.0459 Pa. The authors used two times the single-rotor values and a quadratic fit to obtain the theoretical reference curve in Figure 3. The plotted fit, the values observed by the sensors, and the values recorded by the bell setup thus originate from three described procedures.
The vertical scale of Figure 3 contains repeated printed labels between its lower and upper regions. The delivered figure retains those labels, together with the original symbols. The numerical discussion uses the corresponding tables. Using the tables also keeps the underlying channel labels visible when the same nominal flow is represented by more than one observation. The plot remains useful to distinguish the fitted reference from the measured symbols and to locate the region discussed by the authors.
Nominal flow (m³/h)
Run time (s)
Volume (L)
Pulse quantity N
Frequency (Hz)
Tip speed (m/s)
One-rotor loss (Pa)
100
58.7776
1606.97
154.5448
38.6362
10.8367
193.9085
90
65.0265
1603.33
139.3770
34.8442
9.7732
157.7139
80
73.7457
1599.42
122.5983
30.6496
8.5966
122.0272
70
83.8728
1596.94
107.6282
26.9070
7.5469
94.0459
60
79.3672
1296.30
92.3258
23.0814
6.4739
69.2044
50
91.6543
1295.41
79.8938
19.9734
5.6022
51.8219
40
47.2278
499.06
59.7329
14.9332
4.1885
28.9678
30
61.1878
498.83
46.0836
11.5209
3.2314
17.2417
20
81.6086
448.74
31.0826
7.7707
2.1795
7.8437
15
72.2517
299.28
23.4147
5.8537
1.6418
4.4511
10
54.1953
149.60
15.6037
3.9009
1.0941
1.9767
8
68.3222
149.67
12.3832
3.0958
0.8683
1.2449
5
110.3407
149.73
7.6706
1.9177
0.5379
0.4777
3
185.2778
149.77
4.5694
1.1424
0.3204
0.1695
2
266.0978
149.90
3.1843
0.7961
0.2233
0.0823
1.5
138.1118
49.96
2.0448
0.5112
0.1434
0.0339
1
170.5174
49.96
1.6562
0.4140
0.1161
0.0223
0.65
302.3376
49.96
0.9341
0.2335
0.0655
0.0071
Figure 3. Original theoretical and observed pressure-loss traces and markers across the reported flow range. Source: Zhang, Su and Lin [1], original Figure 3.
Original theoretical and observed pressure-loss traces and markers across the reported flow range.

Results at common flow points

At the nominal 100 cubic metres per hour point, the published setup readings for the bell, nozzle, and master-meter methods are 408, 430 and 435 Pa. The corresponding sensor readings are 364.2, 358.3 and 375.8 Pa. At the nominal 70 cubic metres per hour point, the setup readings are 229, 250 and 220 Pa and the corresponding sensor readings are 191.35, 198.52 and 180.0 Pa. Tables 3 and 4 give the source's stated differences in percentage form and keep both sets of readings with their original labels.
The separation of columns changes how the study's concluding statement should be read. The authors describe the prototype as having a maximum pressure loss of less than or equal to 400 Pa. All three sensor observations in Table 3 fall below that value. All three setup observations in the same table exceed it. A technical comparison quoting this paper can preserve that distinction by identifying which column is being used and the associated nominal flow point. This also permits a buyer to ask whether a new test report uses the same pressure boundaries and channel definition.
The difference between the two columns does not have one fixed value across the three setups. At 100 cubic metres per hour it is smallest in the bell arrangement among these published rows. At 70 cubic metres per hour, the same ordering by the article's relative-difference column holds for these methods. The authors discuss variation and calibration of the equipment and the influence of the laboratory environment as possible explanations for differences in observation. The reported values and that discussion provide a starting point for a comparison under controlled conditions on another instrument.
Test setup
Setup reading (Pa)
Digital sensor reading (Pa)
Relative deviation as published (%)
Bell prover
408
364.2
12.0
Critical-flow Venturi nozzle
430
358.3
20.0
Master-meter system
435
375.8
15.7

Lower-flow results and implications

The low-flow section of Figure 3 shows the observed points moving away from the fitted theoretical reference. The authors discuss the contribution of mechanical and pipe losses to the observed differential in that range. In a low-flow comparison, the expected differential and the performance of the pressure instrumentation are both relevant. A report can retain observations from the chosen flow points and state the applicable resolution, repeat observations and reference setup. This is more informative for that operating range than extending the high-flow comparison by applying a fixed percentage to every point.
For a specification review, the operating-flow basis is also material. A value reported at a nominal laboratory point has a defined meaning when carried with the gas and test state. A pressure-loss requirement for a planned installation should similarly identify the gas state, the flow quantity, and the location of the pressure observations. If a test document gives two channels, both can be kept in the record along with the basis of the reported acceptance or comparison. A change between channels can then be evaluated without changing the meaning of the original results.
The source used a single new rotor design and a limited set of bench arrangements. The published material yields no estimate of the spread of results across production units. If installation evidence is needed, a review can ask for the number of instruments tested, the meter identification and the repeat readings at each agreed point. That record can support a comparison of the assembly that will be supplied.
Test setup
Setup reading (Pa)
Digital sensor reading (Pa)
Relative deviation as published (%)
Bell prover
229
191.35
19.7
Critical-flow Venturi nozzle
250
198.52
25.9
Master-meter system
220
180.0
22.2

Conclusion and limitations

A specification can turn the structure of the source tables into a request for comparable measurements. For each flow point, record the nominal and observed flows, the equipment used to establish that flow, and the instrument used to measure the pressure difference. Preserve the individual observations before reporting a fitted curve or a maximum. If the readings disagree, the paired values tell the reader which quantity changed, while a single reported maximum obscures the pairing. Identify whether a reported limit describes a finite set of observed points or a separate acceptance criterion. Where an acceptance decision is needed, it requires the criterion and measurement uncertainty for that bench and sample.
The study provides original geometry drawings, one comparison plot and numerical data for a gas Roots meter prototype in a laboratory. The data show why the measurement channel and the nominated flow point belong next to every pressure value. The authors' low-flow discussion also supports recording observations near the lower end of the application, where the fitted reference and measured points show different behaviour. The paper gives a basis for specifying what to request from a pressure-loss test. A product-specific determination would use the meter configuration, medium, conditions, pressure boundaries and a documented reference relevant to the planned use.

References and Image Sources

[1] Zhang, G., Su, Z., and Lin, J. (2017). Measurement of pressure loss of a new gas Roots flowmeter. Journal of China University of Metrology, 28(1), 35–39. DOI as printed: 10.3969/j.issn.2096-2835.2017.01.006. Journal issue catalogue
Figures 1–3 and Tables 1–4. Zhang, Su and Lin [1], pp. 36–39. The published rotor drawings and comparison chart are reproduced with English wording on the original chart. Table headings are translated and source observations are transcribed. Copyrighted source figures are used with the authorization confirmed by the publisher of this article.
Cover. Kateryna Babaieva / Pexels. Stock photograph cropped for layout. Photo · Pexels License