Rotary Gas Meters: How Rotor Counts Change the Calibration Window
Abstract
A rotary meter with a low-frequency register output may take many minutes to register enough pulses at minimum flow. Chen and colleagues tested a different acquisition route on a G65 meter. They measured the relationship between rotor motion and the meter's accumulated indication at high flow, then applied that ratio to shorter tests. At the 0.65 m³/h point, the two tabulated indication errors were −0.94% and −0.88%, with acquisition times of 1,098.9174 and 63.458 seconds, respectively. The example illustrates the potential of a rotor signal and the checks needed before transferring the method to another meter and reference bench.
Background and research question
A positive-displacement rotary gas meter traps and releases pockets of gas as two lobed rotors turn. A display or counter can represent that motion at a much lower pulse frequency than the rotors themselves. When a flow bench uses only the counter signal, its measurement window has to accommodate this coarse volume increment. At minimum flow, a long test may still contain only a few pulses.
Chen and colleagues asked whether a measured rotor-to-register ratio could support a shorter indication-error test. Their example is useful because it records pulse counts as well as the final percentage. The number of pulses, the reference volume and the run time show why two acquisition routes can return close errors through different observation windows.
Principle and the two measurement channels
The article distinguishes mechanical meters with a counter, smart meters with a volume corrector, and smart dual-display meters with both outputs. Table 1 preserves the original comparison. Figure 1 shows the succession of rotor positions within one cycle. A gear train transfers rotor motion to the register. The authors distinguish the count-based ratio in that transmission from a nominal swept volume on a nameplate. Small leakage paths and manufacturing variation affect the volume associated with a rotor cycle in an actual flow test.
The proposed bench records a low-frequency counter channel and a rotor channel measured by an optical sensor. The two channels need independent timing so that each is associated with its own reference volume. Table 3 illustrates why that distinction matters: the two 100 m³/h runs list different elapsed times and different reference volumes.
Feature | Mechanical | Smart, single display | Smart, dual display |
Volume measurement | Counter | Volume corrector | Counter + volume corrector |
Signal output | Low-frequency pulse | High- or low-frequency pulse | High- or low-frequency pulse |
Figure 1. Four views of the paired rotors at successive angles, including the 45-degree reference mark and arrows for movement. Original Figure 1 from Chen et al. [1]. Original panel marks and symbols retained.
Equations and bench procedure
For a counter run, the source calculates error from E₁ = [(V₁ − Vₛ₁)/Vₛ₁] × 100% (1). V₁ is the register indication and Vₛ₁ is the reference volume in that interval. The rotor-channel equivalent indicated volume is V₂ = n/K (2), where n is the rotor count in the authors’ notation and K converts the count into volume. With K established, the rotor-channel error is [(n/K − Vₛ₂)/Vₛ₂] × 100% (3). With fractional error e₁ = E₁/100 when E₁ is written as a percentage number, the calibration relation is K = n/[Vₛ₂(1 + e₁)] (4). The reference volume Vₛ₂ belongs to the rotor interval.
The reported specimen was a G65 meter with a flow range of 0.65–100 m³/h. Its counter gave 10 pulses per cubic metre. The source measured paired outputs at 100 m³/h to determine K and listed the same K at the lower points. This procedure assumes that the link between the counting channels is stable during the tests. The tables preserve the authors’ reported values, including the high-flow fields marked with a slash.
Results and interpretation
At 0.65 m³/h the register method recorded two pulses in 1,098.9174 seconds, compared with 36 rotor pulses in 63.458 seconds. Based on the table values, the shorter interval is about 5.8% of the longer one, a reduction by a factor of about 17.3. The paper describes the duration as roughly one twentieth. The error changes from −0.94% to −0.88%, a difference of 0.06 percentage points. At 25 m³/h the two errors were +0.34% and +0.32%, with runs of 85.3384 and 60.256 seconds.
Table 2 describes a counter run at every flow. Table 3 describes high-flow ratio determination followed by shorter rotor runs. Its 100 m³/h counter error belongs to that second sequence. The longer low-flow register interval creates the largest observed difference in acquisition effort. The authors separately report about an hour for the counter-based workflow and about ten minutes for the ratio-based workflow; those are overall process estimates, beyond the acquisition intervals in the tables.
Measurement | 100 m³/h | 25 m³/h | 0.65 m³/h |
Output | Counter | Counter | Counter |
Pulses | 18 | 6 | 2 |
Reference indication (m³) | 1.78727 | 0.59794 | 0.20190 |
Test time (s) | 64.143 | 85.3384 | 1098.9174 |
Meter indication (m³) | 1.8 | 0.6 | 0.2 |
Indication error (%) | 0.71 | 0.34 | −0.94 |
Measurement | 100 m³/h, counter | 100 m³/h, rotor | 25 m³/h, rotor | 0.65 m³/h, rotor |
Output | Counter | Rotor | Rotor | Rotor |
Pulses | 17 | 4896 | 1234 | 36 |
Reference indication (m³) | 1.68783 | 1.67705 | 0.42438 | 0.01253 |
Test time (s) | 60.472 | 60.035 | 60.256 | 63.458 |
Meter indication (m³) | 1.7 | / | 0.42573 | 0.01242 |
Indication error (%) | 0.72 | / | +0.32 | −0.88 |
Transmission ratio K | / | 2898.55 | 2898.55 | 2898.55 |
Conclusion, limitations and follow-up
For this specimen, measuring rotor motion gave an indication-error result close to the register-based result at the two lower points while requiring substantially less time at minimum flow. The paper supports an instrumented bench method for the sample and conditions it documents. The comparison does not establish a general time saving across register designs or a change to acceptance criteria.
A follow-up comparison could measure how much K varies across repeated high-flow determinations, then repeat the low-flow test with independent reference volumes and retain raw timestamps. Useful acceptance evidence would include the reference uncertainty, count handling at interval boundaries and any changes in the signal chain between the two runs.
References and Image Sources
[1] Chen, S., Jin, L., and Wang, H. (2014). Discussion of a new detection technology for gas Roots flowmeters [A new test method for rotary gas meters]. Metrology Technology, 2014(7), 26–28. DOI 10.3969/j.issn.1000-0771.2014.07.08 as printed in the paper.
Official journal issue and article entryFigure 1 and Tables 1–3. Chen et al. [1]. The original schematic has only symbol and panel marks. Original table headers were translated, with the readings and units retained.