Ultrasonic Gas Meters: How Air-to-Gas Shifts Differ Between G4 and G2.5 Samples
Abstract
Shi and colleagues used a piston gas flow standard to compare air and natural-gas indications for six ultrasonic gas meters. The three G4 units showed their largest reported mean shifts at minimum flow, while the three G2.5 units had larger negative shifts across much of the tested range. The two size groups were made by different manufacturers, so the comparison also reflects different sample designs. This article reads the original six sample series and explains the difference between a medium-to-medium shift, an indication error and an error span. It also distinguishes the authors' diagnoses from the measurements themselves.
Background and research question
A flow test in air gives a repeatable way to compare an ultrasonic gas meter with a reference, but it leaves a second question open. Will the same unit produce a similar indication when the working medium is natural gas? An answer requires comparing the same meter at the same flow point. It also helps to check the shape of the error curve because the two media may differ most at one end of the range.
Shi and colleagues tested three G4 meters and three G2.5 meters. This article asks where the largest changes occurred and what the paired results can tell a reader about the air-to-gas relationship. Each size came from a different manufacturer. The size label and the particular sample design are consequently linked in this comparison.
Transit-time principle and the comparison criteria
Figure 1 shows two transducers separated by an acoustic path of length L at angle θ to the pipe axis. In the source model, t_down = L/(c_f + v_m cos θ) (1), and t_up = L/(c_f − v_m cos θ) (2). Here c_f is sound speed in the medium and v_m is the line-averaged gas velocity. Their reciprocal difference gives v_m = [L/(2 cos θ)](1/t_down − 1/t_up) (3). This ideal relationship eliminates the common sound-speed term, while the actual estimate still depends on how the arrival times are identified.
The criteria discussed in the source distinguish the allowed error at a point, the medium-to-medium mean shift and the mean error span in a flow band. Table 1 preserves the criteria as the authors tabulated them for their study. The high band runs from transition flow qₜ through qₘₐₓ. The low band runs from qₘᵢₙ up to qₜ. The reported mean error span is the maximum mean error minus the minimum mean error within the relevant band.
Figure 1. Upstream and downstream time of flight between transducers A and B along an angled acoustic path through the meter. Original Figure 1 from Shi et al. [1]. Chinese labels replaced with English.
Flow region | Maximum permissible error (%) | Maximum mean shift (%) | Mean error span (%) |
qₜ ≤ q ≤ qₘₐₓ | ±1.5 | ±1.5 | ≤2 |
qₘᵢₙ ≤ q < qₜ | ±3.0 | ±3.0 | ≤4 |
Reference and measurement procedure
The authors used a piston gas flow standard with a nominal reference volume of 120 L. They report expanded system uncertainty of 0.25% with k = 2. They compared air and natural gas independently, with ambient temperature of 18–22 °C, ambient pressure of 100–101 kPa and a temperature difference of at most 1 °C between the standard and the meter positions. The G4 sample range was 0.04–6 m³/h and the G2.5 sample range was 0.025–4 m³/h. The six units were listed as accuracy class 1.5.
Their indication equation is E = [(V_i − V_ref)/V_ref] × 100% (4). V_i is the volume inferred from the meter indication, and V_ref is the volume through the meter as established by the reference. At each point the paper used six measurements to calculate a mean. In the tables, the reported medium shift is the gas mean minus the air mean. For example, a change from −0.37% to +0.33% is a shift of +0.70 percentage points. That subtraction uses presented means; it is different from either of the two errors.
Results for the three G4 samples
At the minimum point of 0.04 m³/h, the reported mean shifts were +1.18, +1.58 and +0.98 percentage points for meters 1–3. Other points in Table 2 had smaller shifts in absolute value. The result for the same unit can also change sign across the range. Meter 1 records a shift of −0.62 points at 6.00 m³/h and +1.18 points at 0.04 m³/h. A whole-range average would blur this contrast.
Table 3 shows that the higher-flow gas error span was larger than the corresponding air span for each G4 specimen. The relationship is reversed in the lower band. The paper reports these G4 cases as meeting the criteria it applied. The authors discuss zero offset as a possible reason for the comparatively large shifts at minimum flow, where the flow-related timing difference is small. The error pattern gives a reason to investigate the zero offset in a new test. The six-measurement means by themselves do not independently establish its magnitude.
Flow (m³/h) | Air mean (%) No. 1 | Air mean (%) No. 2 | Air mean (%) No. 3 | Gas mean (%) No. 1 | Gas mean (%) No. 2 | Gas mean (%) No. 3 | Shift (%) No. 1 | Shift (%) No. 2 | Shift (%) No. 3 |
6.00 | −0.18 | −0.07 | −0.27 | −0.80 | −0.76 | −0.78 | −0.62 | −0.69 | −0.51 |
4.20 | −0.18 | −0.06 | −0.15 | −0.49 | −0.36 | −0.43 | −0.31 | −0.30 | −0.28 |
2.40 | 0.09 | 0.48 | −0.08 | 0.16 | 0.70 | −0.03 | 0.07 | 0.22 | 0.05 |
1.20 | −0.03 | −0.06 | −0.13 | 0.42 | 0.57 | 0.40 | 0.45 | 0.63 | 0.53 |
0.60 | −0.20 | −0.37 | −0.25 | 0.28 | 0.33 | 0.23 | 0.48 | 0.70 | 0.48 |
0.40 | −0.29 | −0.49 | −0.21 | 0.03 | 0.04 | 0.22 | 0.32 | 0.53 | 0.43 |
0.20 | −0.03 | −0.12 | 0.29 | 0.09 | 0.28 | 0.49 | 0.12 | 0.40 | 0.20 |
0.12 | −0.06 | −0.08 | 0.00 | 0.15 | 0.45 | 0.44 | 0.21 | 0.53 | 0.44 |
0.04 | −1.39 | −1.04 | −1.03 | −0.21 | 0.54 | −0.05 | 1.18 | 1.58 | 0.98 |
Meter | Flow region | Air span (%) | Gas span (%) | Maximum mean shift (%) |
1 | High | 0.29 | 1.22 | −0.62 |
1 | Low | 1.36 | 0.36 | 1.18 |
2 | High | 0.85 | 1.46 | 0.70 |
2 | Low | 0.96 | 0.50 | 1.58 |
3 | High | 0.19 | 1.18 | 0.53 |
3 | Low | 1.32 | 0.54 | 0.98 |
Results for the three G2.5 samples
The G2.5 data show negative shifts at most tabulated points. At 0.400 m³/h the reported shifts were −1.74, −2.47 and −2.07 points. Table 5 shows that the maximum reported high-band shifts for all three G2.5 samples exceeded the study’s ±1.5% criterion in magnitude. The low end is not uniform, however. At 0.025 m³/h, sample 5 has a positive shift of +1.92 points, while samples 4 and 6 have negative shifts. The data therefore support a pattern across most points, alongside a specific exception.
The authors relate the broader negative pattern to timing picks that may select the wrong waveform cycle. They give E′ = [(t_up − t_p)(t_down − t_p) − t_up t_down]/(t_up t_down) × 100% (5), with t_p as the acoustic period, as a worked illustration of that kind of timing error. Their example at 500 kHz uses t_p = 2 μs and yields about −1.59%. That calculation is separate from the measured sample means. No independently measured timing trace is presented to confirm the same cause for each row.
Flow (m³/h) | Air mean (%) No. 4 | Air mean (%) No. 5 | Air mean (%) No. 6 | Gas mean (%) No. 4 | Gas mean (%) No. 5 | Gas mean (%) No. 6 | Shift (%) No. 4 | Shift (%) No. 5 | Shift (%) No. 6 |
4.000 | −0.15 | −0.17 | −0.37 | −0.95 | −1.82 | −1.75 | −0.80 | −1.65 | −1.38 |
2.800 | −0.28 | −0.32 | −0.22 | −1.13 | −2.01 | −1.88 | −0.85 | −1.69 | −1.66 |
1.600 | −0.26 | −0.39 | −0.07 | −1.08 | −1.76 | −1.90 | −0.82 | −1.37 | −1.83 |
0.800 | −0.05 | −0.21 | 0.06 | −1.86 | −2.54 | −1.92 | −1.81 | −2.33 | −1.98 |
0.400 | −0.23 | −0.22 | 0.41 | −1.97 | −2.69 | −1.66 | −1.74 | −2.47 | −2.07 |
0.250 | −0.53 | −0.26 | 1.16 | −2.37 | −2.72 | −1.01 | −1.84 | −2.46 | −2.17 |
0.125 | −0.32 | −0.90 | 1.66 | −2.08 | −2.93 | −0.48 | −1.76 | −2.03 | −2.14 |
0.075 | −0.09 | −1.47 | 1.68 | −1.95 | −2.82 | −0.29 | −1.86 | −1.35 | −1.97 |
0.025 | 0.42 | −1.45 | 1.14 | −0.30 | 0.47 | 0.61 | −0.72 | 1.92 | −0.53 |
Meter | Flow region | Air span (%) | Gas span (%) | Maximum mean shift (%) |
4 | High | 0.23 | 1.02 | −1.81 |
4 | Low | 0.95 | 2.07 | −1.86 |
5 | High | Results for the three G2.5 samples | 0.93 | −2.47 |
5 | Low | 1.21 | 3.40 | −2.46 |
6 | High | 0.78 | 0.26 | −2.07 |
6 | Low | 0.54 | 1.62 | −2.17 |
Conclusion, limitations and follow-up
The strongest contrast in this dataset is where the medium change appears on the curve. The G4 changes were largest at minimum flow, whereas the G2.5 cases showed broad negative shifts with an exception at the lowest point. Because the meter models and manufacturers differ, the study cannot isolate meter size as the cause. A follow-up could test more models within each size and retain the zero and waveform measurements, as well as the indicated volumes.
For a model comparison, ask for air and gas measurements at matched points, the test uncertainty and the full error tables. Retain raw indication errors, the band spans and the signed medium difference as separate fields. This lets a reader distinguish a change in the entire error curve from a localized effect.
References and Image Sources
[1] Shi, H., Xiang, L., Jin, L., Chen, S., Chen, C., and Zhang, Q. (2024). Characterization Study of Gas-Air Relationship Measurement of Ultrasonic Gas Meter. Process Automation Instrumentation, 45(7), 11–15.
Bibliographic cross-reference, ref. 3 ·
Independent bibliographic cross-reference, ref. 2Figure 1 and Tables 1–5. Shi et al. [1]. Original transit-time diagram with English transducer labels. Tables retain the source numbers, units and grouping.